Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
China Occupational Medicine ; (6): 594-596, 2016.
Article in Chinese | WPRIM | ID: wpr-876996

ABSTRACT

OBJECTIVE: To establish a methodology to dilute blood sample with compound matrix modifier,and to determine the blood level of nickel using improved graphite furnace atomic absorption spectrometry( GFAAS). METHODS: The blood samples( 100. 00 μL) were diluted 5 times in 0. 12 g / L mass concentration of palladium chloride combined with 1. 00 %(V /V) Triton X-100 and 0. 10%(V /V) nitric acid,and the blood level of nickel was detected by GFAAS. RESULTS: The good linearity range of nickel mass concentration in blood was 0. 30-25. 00 μg / L,the correlation coefficient was 0. 999 7,the detection limit was 0. 30 μg / L,the minimum detectable mass concentration was 1. 50 μg / L,the recovery rate was99. 00 %-102. 75 %. The within-run and between-run relative standard deviations were 1. 43 %-3. 70 % and 2. 20 %-5. 03 %,respectively. Blood samples can be stored for at least 7 days at 4 ℃. CONCLUSION: This method is simple for operation,with high accuracy and precision,which is suitable for determination of nickel in blood in normal individuals and workers with occupational exposed to nickel.

SELECTION OF CITATIONS
SEARCH DETAIL